units mismatch in resolution calculation

Reported by: brian.maranville@… Owned by: srkline
Priority: minor Milestone:
Component: SANS Reduction Keywords:
Cc: Blocking:
Task:

Description

It seems like line 136 (and line 284) in version 7.94 of NCNR_User_Procedures/Reduction/SANS/NCNR_utils.ipf have a unit mismatch - (applies to resolution calculation when lenses present)

v_b = 0.25*(S1*L2/L1)^2 +0.25*(2/3)*(lambdaWidth/lambda)^2*(S2*L2/lp)^2		//correction to 2nd term

because S1, L2 and L1 are in cm, the first term has units of cm2, which I would expect, but the second term has units of cm2/Angstroms, because lambdaWidth is unitless (it is delta_lambda/lambda) but lambda is in Angstroms, and S2, L2 and lp are in cm.